Senseca ECO 130 handleiding

20 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 9 van 20
ECO130 - Thermocouple Quick Response Thermometer
Operation
B-H89.0.0X.DK2-2.1 Page 9 of 20
4 Operation
4.1 Opening the configuration menu
1 Press the Function button for 2 seconds to open the Configuration menu.
2 (ONF appears in the display. Release the Function button.
Parameter
Values
Meaning
POFF
Shut-off time
OFF
No automatic shut-off
0:15, 0:30, 1:00, 4:00,
12:00
Automatic shut-off after a selected time in
hours:minutes, during which no buttons have been
pressed
L,TE
Backlight
OFF
Backlight deactivated
0:15, 0:30, 1:00, 2:00,
4:00
Automatic shut-off of the backlight after a selected
time in minutes:seconds, during which no buttons
have been pressed
ON
No automatic shut off of the backlight
UN,T
Display unit
°(
Temperature display in °C
°F
Temperature display in °F
IN,T
Factory settings
NO
Use current configuration
YES
Reset device to factory settings.
After confirming with the function-button, the
display shows: IN,T DONE
4.2 Adjustment of the measuring input
The temperature input can be adjusted with the zero point correction and the gradient cor-
rection. If an adjustment is made, you change the pre-adjusted factory settings. This is
signaled with the display text T.OF or T.SL when switching on.