Senseca ECO 130.2 handleiding
Handleiding
Je bekijkt pagina 9 van 22

ECO 130.2 - Quick response 2 channel-alarm-thermometer
Operation
B-H89.0.1X.DK2-2.1 Page 9 of 22
4 Operation
4.1 Opening the configuration menu
1 Press the Function button for 2 seconds to open the Configuration menu.
2 (ONF appears in the display. Release the Function button.
Parameter
Values
Meaning
INP
Measuring channel
(H.1
Main display = channel 1
BOTH
Main display = channel 1
Secondary display = channel 2
D,FF
Main display = difference channel 1 – channel 2
Secondary display = channel 2
AL.
Alarm *
1
OFF
No alarm active
ON
Alarm via text insertion, acoustic signal and flashing
of the background lighting
BEEP
Alarm via text insertion and acoustic signal
L,TE
Alarm via text insertion and flashing of the back-
ground lighting
AL.LO
Min. alarm limit (only available if AL <> off)
-65.0 .. AL.X,
a min. alert is triggered if the value falls below this
value. (for °F: -85.0 .. AL.Hi)
AL.XI
Max. alarm limit (only available if AL <> off)
AL.LO .. 1200.0
If the value is exceeded, a max. alarm is triggered.
(for °F: AL.Hi .. 2192)
POFF
Shut-off time
OFF
No automatic shut-off
0:15, 0:30, 1:00, 4:00,
12:00
Automatic shut-off after a selected time in
hours:minutes, during which no buttons have been
pressed
L,TE
Backlight