Senseca ECO 121 handleiding

22 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 11 van 22
ECO 120 / ECO 121 - Alarm thermometer
Operation
B-H86.0.0X.DK2-4.1 Page 11 of 22
Parameter
Values
Meaning
L,TE
Backlight
OFF
Backlight deactivated
15, 30, 60, 120, 240
Automatic shut-off of the backlight after a select-ed
time in seconds, during which no buttons have been
pressed
ON
No automatic shut off of the backlight
UN,T
Display unit
°(
Temperature display in °C
°F
Temperature display in °F
IN,T
Factory settings
NO
Use current configuration
YES
Reset device to factory settings.
After confirming with the function-button, the
display shows: IN,T DONE
4.2 Adjustment of the measuring input
The temperature input can be adjusted with the zero point correction and the gradient cor-
rection. If an adjustment is made, you change the pre-adjusted factory settings.
This is signaled with the display text T.OF or T.SL when switching on.
1 Switch the device off.
2 Hold the down button and press the On/Off button briefly to switch on the device and
open the Adjustment menu.
3 The display shows the first parameter. Release the down button.
Parameter
Values
Meaning
T.OF
Zero point correction
0.00
No zero point correction
-5.00 5.00
Zero point correction in °C. (at °F -9.00 .. 9.00)

Bekijk gratis de handleiding van Senseca ECO 121, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.

Productinformatie

MerkSenseca
ModelECO 121
CategorieNiet gecategoriseerd
TaalNederlands
Grootte2193 MB