ECO 121-H - HACCP alarm thermometer
Error and system messages
B-H86.0.0Y.DK2-4.1 Page 17 of 22
7 Error and system messages
Display
Meaning
Possible causes
Remedy
----
Measurement far
outside of the
measuring range
Sensor cable or probe
defect
• Measurement
outside of the
measuring range
• Probe or device
defect
• Measuring probe or
device defect
Stay within allowable
measurement range
Send in for repair
Send in for repair or
connect a different
probe
N
o display,
unclear char-
acters or no
response
when buttons
are pressed
Battery depleted
System error
Device is defective
• Battery depleted
• Error in the device
Replace battery
Send in for repair
BAT
Battery depleted
• Battery depleted
Replace battery
ERR.1
Measuring range
exceeded
• Measurement too
high
• Measuring probe or
device defect
Stay within allowable
measurement range
Send in for repair
ERR.2
Measuring range is
undercut
• Measurement too
low
• Measuring probe or
device defect
Stay within allowable
measurement range
Send in for repair
SYS ERR
System error
• Error in the device
Switch device on/off
Replace batteries
Send in for repair