Senseca ECO 121-H handleiding

22 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 17 van 22
ECO 121-H - HACCP alarm thermometer
Error and system messages
B-H86.0.0Y.DK2-4.1 Page 17 of 22
7 Error and system messages
Display
Meaning
Possible causes
Remedy
----
Measurement far
outside of the
measuring range
Sensor cable or probe
defect
Measurement
outside of the
measuring range
Probe or device
defect
Measuring probe or
device defect
Stay within allowable
measurement range
Send in for repair
Send in for repair or
connect a different
probe
N
o display,
unclear char-
acters or no
response
when buttons
are pressed
Battery depleted
System error
Device is defective
Battery depleted
Error in the device
Replace battery
Send in for repair
BAT
Battery depleted
Battery depleted
Replace battery
ERR.1
Measuring range
exceeded
Measurement too
high
Measuring probe or
device defect
Stay within allowable
measurement range
Send in for repair
ERR.2
Measuring range is
undercut
Measurement too
low
Measuring probe or
device defect
Stay within allowable
measurement range
Send in for repair
SYS ERR
System error
Error in the device
Switch device on/off
Replace batteries
Send in for repair

Bekijk gratis de handleiding van Senseca ECO 121-H, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.

Productinformatie

MerkSenseca
ModelECO 121-H
CategorieNiet gecategoriseerd
TaalNederlands
Grootte2225 MB