ECO 121-H - HACCP alarm thermometer
Operation
B-H86.0.0Y.DK2-4.1 Page 11 of 22
Parameter
Values
Meaning
POFF
Shut-off time
OFF
No automatic shut-off
15, 30, 60, 120, 240
Automatic shut-off after a selected time in minutes,
during which no buttons have been pressed
L,TE
Backlight
OFF
Backlight deactivated
15, 30, 60, 120, 240
Automatic shut-off of the backlight after a select-ed
time in seconds, during which no buttons have been
pressed
ON
No automatic shut off of the backlight
UN,T
Display unit
°(
Temperature display in °C
°F
Temperature display in °F
IN,T
Factory settings
NO
Use current configuration
YES
Reset device to factory settings.
After confirming with the function-button, the
display shows: IN,T DONE
4.2 Adjustment of the measuring input
The temperature input can be adjusted with the zero point correction and the gradient cor-
rection. If an adjustment is made, you change the pre-adjusted factory settings.
This is signaled with the display text T.OF or T.SL when switching on.
1 Switch the device off.
2 Hold the down button and press the On/Off button briefly to switch on the device and
open the Adjustment menu.
3 The display shows the first parameter. Release the down button.