4 Operation
4.1 Opening the configuration menu
1 Press the Function key for 2 seconds to open the Configuration menu.
2 (ONF appears in the display. Release the Function key.
Parameter
Values
Meaning
AXLD
Auto Hold-Function (factory setting: oFF)
OFF
Auto Hold deactivated: Value is frozen on keypress
(hold)
ON
Auto Hold activated: automatic holding of the
measured value, as soon as it is stabilized.
CO.DO
Count-Down activated: Freeze value after defined
time
CO.DO
Count-Down time in seconds (only available if AHLd = co.do)
5…240
Count-Down time in seconds, after which
temperature is frozen in display automatically (start
with OK) (factory setting: 30s)
AL.
Alarm
OFF
No alarm active
ON
Alarm via text insertion, acoustic signal and flashing
of the background lighting
BEEP
Alarm via text insertion and acoustic signal
L,TE
Alarm via text insertion and flashing of the back-
ground lighting
AL.LO
Min. alarm limit (only available if AL <> off)
-70.0 .. AL.X,
a min. alert is triggered if the value falls below this
value. (at °F: -94.0 .. AL.Hi)
AL.XI
Max. alarm limit (only available if AL <> off)
AL.LO .. 250.0
If the value is exceeded, a max. alarm is triggered.
(at °F: AL.Hi .. 482.0)