2. About broncolor Radio System
●
7.
Press the Measuring Button 6.
The ash will re and the measured value (F-stop) will be displayed.
All ashes in the same Studio Address will re even if a particular Lamp Address is
selected.
NOTICE
In case of the following, please follow “Cord Flash Mode” ( ).
● When ring the ash, if the ash brightness is lower than the ambient light, the meter
may fail to detect the light.
● Rapid start uorescent lamps and special lighting are sometimes mistaken for ash and
accidentally measured.
● Even if the ash does not re, when a sudden light change occurs in the light receptor,
measurement may be made.
● The waveform of a ash bulb has a slight slope and there is a possibility that the light
meter cannot recognize the ash bulb.
NOTE
● With broncolor radio system, individual ash triggering is not available for each Lamp
Address. All ashes in the same Studio Address will re even if a particular Lamp
Address is selected. For measuring the ash individually, turn off the power of ash units
not in use. Or use the retracted lumisphere of L-858D to measure the light in front while
all ashes re.
● For red-eye prevention and auto-ash light adjustment, some devices can pre-ash
before main ash burst. With the normal setting, the light meter will measure the pre-
ash bursts and not the main ash burst. To take a successful reading, activate the pre-
ash feature in the Tool Box. Please refer to the L-858D Operating Manual.
● For radio frequencies, refer to “5. Radio Studio Address Frequencies”.(
P36)
[Measuring Screen]
Measured Value
(F-stop)