Sauermann Si-CPE320 handleiding

38 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 34 van 38
34 Modbus
4280 BOOLEAN Audible alarm 0: no sound, 1: Sound in case of alarm
4290 U8 Acknowledgement duration Allowed delay to acknowledge the
alarm.
This value must belong to range [0,
60]
13.5.5 Probes and module parameters
Modbus Register type Description Possibilities
Probe Entry 1
6000 F32 Lower limit of probe range, for
each available measure
For maximum 10 measures (@6000:
lower range of measure 1, @6002:
lower range of measure 2, etc…).
6020 F32 Upper limit of probe range, for
each available measure
For maximum 10 measures (@6000:
upper range of measure 1, @6002:
upper range of measure 2, etc…).
6040 U8 Atmospheric pressure compen-
sation unit
cf. units enumeration
6050 F32 Atmospheric pressure compen-
sation value. In case of pressure,
this value is used for velocity cal-
culation
In Pa.
6060 U8 Type of pressure equipment
plugged in order to determine
the air velocity
0: Pitot tube S, 1: Pitot tube L, 2: de-
bimo blades, 3: other (set pressure
Equipment Coeff instead), 4: None
6070 F32 Custom coefcient for pressure
device plugged in order to deter-
mine the air velocity
From 0.0001 to 9.9999.
6080 U8 Room volume unit cf. units enumeration
6090 F32 Room volume value in cubic
meter used for computing air re-
newing rate (ACR)
In cubic meter.
6100 U8 Measurement integration From 0 to 9.
6110 U8 Temperature compensation
mode
0: Use internal sensor, 1: Manual – re-
fer to user value, 2: Use probe id 0 (or
user if not present), 3: Use probe id 1
(or user if not present), 4: Use probe
id 2 (or user if not present).
6120 U8 Manual temperature compensa-
tion unit
cf. units enumeration
6130 F32 Manual temperature compensa-
tion value in Celsius degrees
From -50 to 50, in Celsius degrees.
6140 F32 Correction factor used for air ve-
locity and airow probes
From 0.2 to 2 - Used for air velocity
and airow probes (default = 1).
6150 U8 Type of section used for air ve-
locity and airow probes
0: rectangular, 1: circular, 2: other (set
airow coefcient instead), 3: None
(not congured)
6160 U8 Unit for section diameter, section
length and section width
cf. units enumeration
6170 F32 Section diameter value in meters
(used when section type is circu-
lar)
From 0.001 to 3.