Measurement Threshold
• First, click or tap % or Abs as the display type.
• Click or tap the drop-down button of Source to select the desired channel. The
available channels include analog channels and Math1-Math4.
• Click or tap the input field of Upper to set the upper limit of the measurement
with the pop-up numeric keypad. You can also use the specified knob to set it.
When the upper limit is set to be smaller than or equal to the current middle
value, a prompt message "Set at lower limit" is displayed. Then, the oscilloscope
will automatically adjust the upper limit and make it greater than the middle
value. By default, it is 90%. The default absolute value varies with the vertical
setting of the channel.
• Click or tap the input field of
Mid to set the middle value of the measurement
with the pop-up numeric keypad. You can also use the specified knob to set it.
The middle value is limited by the settings of the upper limit and lower limit. By
default, it is 50%. The default absolute value varies with the vertical setting of the
channel.
• Click or tap the input field of
Lower to set the lower limit of the measurement
with the pop-up numeric keypad. You can also use the specified knob to set it.
When the lower limit is set to be greater than or equal to the current middle
value, a prompt message "Set at upper limit" is displayed. Then, the oscilloscope
will automatically adjust the lower limit and make it smaller than the middle
value. By default, it is 10%. The default absolute value varies with the vertical
setting of the channel.
• Click or tap
Default, and then the upper value, middle value, and lower value will
be restored to the defaults.
Click or tap the ON/OFF tab for
Threshold to enable or disable the threshold settings.
TIP
Modifying the threshold will affect the measurement results of time, delay, and phase
parameters.
Measurement Range
Click or tap the drop-down button of
Region to select "Main" or "Zoom".
• Main: indicates that the measurement range is within the main time base region.
• Zoom: indicates that the measurement range is within the zoomed time base
region.
"Zoom" is available to choose after you have enabled
Zoom Mode (Delayed
Sweep)
first.
• Cursor: When CursorAB is set to ON, then two cursors will be displayed on the
screen. At this time, click or tap the input field of
CursorA and CursorB
Measurements
MHO2000 User Guide
154
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.