M300 Quick Guide
26
To Configure Scan List
Create a scan list and add any channels of the multiplexer or the DIO or TOT
channels of the multifunction module to the scan list; then, configure the basic
measurement, scaling, alarm and advanced measurement parameters for each
channel. The edited scan list can be stored in internal or external memory and be
recalled when required.
1 Create a new list
Press Config New, input the desired list name and press OK.
Note: The existing scan list will be overwritten by the scan list currently created.
2 Set the scan parameters
(1) Select Sweep and use the left/right arrow key to select “Infinite” or “1”
(you need to set a value and the range is from 1 to 50000).
(2) Select Trigger Mode and use the left/right arrow key to select “Auto”,
“Manual”, “External”, “AbsTime” or “Alarm”.
(3) If “Auto” is selected, please select Interval and use the numeric keyboard
to set a desired interval. The range is from 00:00:00.000 to 99:59:59.999
(the time format is Hour:Minute:Second.Millisecond) and the default is 0.
If “External” is selected, please select Edge and use the left/right arrow key
to select “Rising” or “Falling”.
If “AbsTime” is selected, please select AbsTime and use the numeric
keyboard to set a desired time. The time format is “Month-Date
Hour:Minute:Second”.
If “Alarm” is selected, please select Channel and use the left/right arrow
key to select “Alarm1”, “Alarm2”, “Alarm3” or “Alarm4”.
3 Configure a channel to the scan list (taking the DCV measurement
function as an example
)
Press ConfigEditNew to enter the channel configuration guide.
(1) Step 1: Measurement Configuration
Select Chan No. and use the left/right arrow key to select the desired
channel number (for example, 101);
Select Function and use the left/right arrow key to select the desired
function (for example, DCV);
Select Range and use the left/right arrow key to select the desired range
(for example, 2V).
Note: The parameters need to be set in this step depend on the
measurement function selected.
(2) Step 2: Scaling Configuration
Select Scaling and use the left/right arrow key to enable the scaling