RaySafe X2 MAS
Stop delay setting shorter than the time between pulses (t<T), will result in many short measure-
ments (one for each pulse).
SETTING: IGNORE PREPULSES
Use the ignore prepulses setting to remove one or many unwanted prepulses from the measure-
ment.
Ignore prepulses = 0, the whole exposure is captured.
Ignore prepulses = 1, the (first) prepulse is excluded from the measurement.
Note! This setting will affect all parameters, including the dose measurement.
SETTING: MA MODE
Select mA avg when you want to measure the average tube current on a pulsed exposure. Other-
wise, use mA.