RaySafe X2 handleiding
Handleiding
Je bekijkt pagina 50 van 51

RaySafe X2 – MAS
Stop delay setting shorter than the time between pulses (t<T), will result in many short measure-
ments (one for each pulse).
SETTING: IGNORE PREPULSES
Use the ignore prepulses setting to remove one or many unwanted prepulses from the measure-
ment.
Ignore prepulses = 0, the whole exposure is captured.
Ignore prepulses = 1, the (first) prepulse is excluded from the measurement.
Note! This setting will affect all parameters, including the dose measurement.
SETTING: MA MODE
Select mA avg when you want to measure the average tube current on a pulsed exposure. Other-
wise, use mA.
49
Bekijk gratis de handleiding van RaySafe X2, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.
Productinformatie
| Merk | RaySafe |
| Model | X2 |
| Categorie | Niet gecategoriseerd |
| Taal | Nederlands |
| Grootte | 6337 MB |







