RaySafe X2 MAS
Switch on the X-ray generator.
Make an exposure and read the result.
Note! Users of RaySafe X2 mAs must be aware of the potential damage to generators and of
electrical human hazards in case of improper connection or damaged equipment. RaySafe X2
mAs is intended for use only by personnel authorized in performing calibration and repair of X-ray
equipment.
Tip! You can measure with the R/F, DENT or MAM Sensor and the mAs cable both connected
to get radiation and generator current values at the same time.
Tip! Tap a parameter for large digits, to get information about that specific parameter, and, if
applicable, see a waveform.
MAS: MEASUREMENT PARAMETER DEFINITIONS
mAs is calculated from all recorded data.
mA is calculated as an average of all samples above 50% of peak. Rush currents are automati-
cally removed. Live readings are based on samples since last reading. For long measurements, the
final reading is recorded approximately 1–2 s before end trig.
mA avg (average) is calculated as mAs / time. For measurements longer than 3 s, the final read-
ing is a moving average ending approximately 1–2 s before end trig. Live readings are based on a
1 s moving average. (Change mA mode to mA avg to measure according to this definition.)
Time starts the first time the current reaches 50% of the calculated mA and ends the last time it
drops below 50%. Live readings are time since start trig.