RaySafe X2 R/F
Ignore prepulses = 1, the (first) prepulse is excluded from the measurement.
Note! This setting will affect all parameters, including the dose measurement.
SETTING: AC PULSE COUNTING
The AC pulse counting method can help in certain situations, but should normally be turned Off.
Exceptions:
To avoid counting pre-heat pulses on intra oral AC machines, select On.
To count individual pulses on full wave rectified single phase AC machines, select On.
Note! This setting affects pulses, pulse rate and dose per pulse.
SETTING: CORRECTION FACTOR
For some specific measurement setups, a correction factor is applicable. In all other cases, use
None.
Siemens CT Straton
Some Siemens CT Somatom machines have a Straton tube. To get a correct kV value on such
machines, select Siemens CT Straton. This setting affects only the kVp value.
GE CT 10.5
°
When measuring kV on a GE CT machine with 10.5
°
anode angle, select GE CT 10.5
°
to use
a correction factor specifically adapted for maximum kV accuracy on such machines. Measure
without bow tie filter if possible. This setting affects only the kVp value and is intended for GE
Field Service Engineers.
GE CT 7
°
Monopolar
When measuring kV on a GE CT machine with 7
°
anode angle and monopolar tube, select GE
CT 7
°
Monopolar to use a correction factor specifically adapted for maximum kV accuracy on
such machines. Measure without bow tie filter if possible. This setting affects only the kVp value
and is intended for GE Field Service Engineers.
16