RaySafe X2 handleiding

51 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 16 van 51
RaySafe X2 R/F
Use a longer stop delay when measuring on a pulsed fluoroscopy system or to get a prepulse recorded
into the same measurement as the following ordinary exposure.
Stop delay setting longer than the time between pulses (t>T), will result in one long measure-
ment. Note that you have to wait as long as the stop delay setting (t) after the last exposure be-
fore the measurement ends and values will be displayed.
Stop delay setting shorter than the time between pulses (t<T), will result in many short measure-
ments (one for each pulse).
SETTING: IGNORE PREPULSES
Use the ignore prepulses setting to remove one or many unwanted prepulses from the measure-
ment.
Ignore prepulses = 0, the whole exposure is captured.
15

Bekijk gratis de handleiding van RaySafe X2, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.

Productinformatie

MerkRaySafe
ModelX2
CategorieNiet gecategoriseerd
TaalNederlands
Grootte6337 MB