RaySafe X2 – R/F
Use a longer stop delay when measuring on a pulsed fluoroscopy system or to get a prepulse recorded
into the same measurement as the following ordinary exposure.
Stop delay setting longer than the time between pulses (t>T), will result in one long measure-
ment. Note that you have to wait as long as the stop delay setting (t) after the last exposure be-
fore the measurement ends and values will be displayed.
Stop delay setting shorter than the time between pulses (t<T), will result in many short measure-
ments (one for each pulse).
SETTING: IGNORE PREPULSES
Use the ignore prepulses setting to remove one or many unwanted prepulses from the measure-
ment.
Ignore prepulses = 0, the whole exposure is captured.
15