7
PCE Instruments | www.pce-instruments.com
HOW TO ENTER THE ‘FORCE MENU
With the instrument in normal modeof operation (see section 10.2),
press and hold the UP(5) key for 1 second. The horizontal leds light
from bottom to top. When the upper led lights, the instrument enters into
the ‘force’ menu.
If the ‘UP(5) key is released before entering into the ‘forcemenu, the
horizontal leds light downwards from top to bottom, and the instrument
returns to ‘normal mode’ of operation.
HOW TO OPERATE INSIDE THE ‘FORCE’ MENU
The available functions inside the forcemenu can be congured (see
section 13.6). By default, Force High’, Force Low and Force Set’ are
available. Inside the ‘force’ menu:
press the ‘UP’ (5) key to move to the next function.
press the ‘SQ’ (<) key to activate the selected function.
When the function is active, the display will remain ashing. Press the
SQ(<) key to deactivate the function (display stops ashing), or wait
for the rollback to activate.
DESCRIPTION OFFORCE’ FUNCTIONS
The ‘forcefunctions allow to manually force the output signal to the low
and high levels of the output signal selected. These functions allow to
easily validate the correct function of remote elements connected to the
instrument output, such as PLC, HMI’s, SCADAs, etc.
The force lowfunction sets the output signal to the minimum value of
the selected range (4 mA or 0 Vdc or the value congured at the ‘output_
low’ parameter).
The force highfunction sets the output signal to the maximum value
of the selected range (20 mA or 10 Vdc or the value congured at the
output_high’ parameter).
The force setfunction sets the output signal to a value between 0 and
100% of the maximum selected range (4 to 20 mA or 0 to 10 Vdc or
the range congured at the output_lowand output_highparameters).
When entering the force set’ function, the display reads ‘50’ (the output
is forced to 50% of the congured range). Use keys UP (5) and ‘LE’ (3)
to move up to 100% or down to 0% of the congured range.
10.4 How to operate the ‘Force’ menu
HOW TO ACTIVATE ‘MESSAGES’ FUNCTION
With the instrument in normal modeof operation (see section 10.2),
press the LE (3) key to activate the messagesfunction. The messages
function displays information about the instrument. The information
available is congurable (see section 13.7).
The messages function ends when all the information has been
displayed or front keys UP(5) or SQ(<) are pressed. The display
returns to ‘normal mode’ of operation.
10.5 How to activate the ‘Messages’ function
10. How to operate the instrument (cont.)
Table 7 | Example of ‘Force’ menu with all functions set to ‘on
‘Force Low’
‘Force high
Exit
Force Set
See section 13.6 for a list and a description of available functions.
10.6 Fast and advanced congurations
FAST CONFIGURATION
The fastest way to congure the instrument is to activate one of the
predened conguration codes (see section 8).
Access the conguration menu and enter the Function code (codE)
menu entry. The code displayed is the current active input - output range.
Select the new code and validate. Selecting a code automatically exits
the ‘conguration menu’ and activates the new conguration.
*
There are different codes for 4/20 mA and 0/10 Vdc output
signals.
To customize the input and output signals, see the Advanced scaling
section of the ‘conguration menu’ (see section 13.4).
ADVANCED CONFIGURATION
Additional conguration parameters are available at the conguration
menu’. The operator can customize the input and output signal ranges,
the message seen on display, the functions available at the ‘forcemenu,
the messages associated to the LE(3) key, activate lters, password
function, etc.
See section 13 for a detailed explanation on the ‘conguration menu’.
All Vdc and Adc input signal ranges can be customized to read
bipolar ranges. Check the Advanced scaling’ section of the
conguration menu’ (see section 13.4).
HOW TO EXIT ‘FORCE’ MENU
To exit the forcemenu, press the LE(3) key, or press the key UP
(5) key until the parameter ---’ appears, and select by pressing the SQ
(<) key, or wait without pressing any key until the automatic rollback
activates.
When exiting the ‘force’ menu, the horizontal leds light down from top to
bottom, and the instrument returns to ‘normal mode’ of operation.
‘ROLLBACK’ FUNCTION
If there is no interaction from the operator for 60 seconds, the instrument
exits the ‘force’ menu and returns to ‘normal mode’ of operation.