PCE Instruments PCE-SCI-E handleiding
Handleiding
Je bekijkt pagina 4 van 24

4
PCE Instruments
Table 2 | Predened conguration codes - Input / Output
Input Signal
Range
Output 4/20 mA
Code
Output 0/10 Vdc
Code
See section ...
0/600 Vac 010 110
11.1
0/450 Vac 011 111
0/300 Vac 012 112
0/150 Vac 013 113
0/100 Vac 014 114
0/60 Vac 015 115
0/30 Vac 016 116
0/15 Vac 017 117
0/10 Vac 018 118
0/2 Vac 019 119
0/1 Vac 020 120
0/500 mVac 021 121
0/300 mVac 022 122
0/200 mVac 023 123
0/150 mVac 024 124
0/100 mVac 025 125
0/75 mVac 026 126
0/60 mVac 027 127
0/50 mVac 028 128
Reserved 029 to 031 129 to 131
0/600 Vdc 032 132
11.2
0/450 Vdc 033 133
0/300 Vdc 034 134
0/150 Vdc 035 135
0/100 Vdc 036 136
0/60 Vdc 037 137
0/30 Vdc 038 138
0/15 Vdc 039 139
0/10 Vdc 040 140
0/2 Vdc 041 141
0/1 Vdc 042 142
0/500 mVdc 043 143
0/300 mVdc 044 144
0/200 mVdc 045 145
0/150 mVdc 046 146
0/100 mVdc 047 147
0/75 mVdc 048 148
0/60 mVdc 049 149
0/50 mVdc 050 150
Reserved 051 to 054 151 to 154
8. Predened conguration codes
Table 2 | Predened conguration codes - Input / Output
Input Signal
Range
Output 4/20 mA
Code
Output 0/10 Vdc
Code
See section ...
0/5 Aac 055 155
11.3
0/4 Aac 056 156
0/3 Aac 057 157
0/2 Aac 058 158
0/1 Aac 059 159
0/500 mAac 060 160
0/300 mAac 061 161
0/75 mAac 062 162
0/50 mAac 063 163
0/20 mAac 064 164
0/10 mAac 065 165
0/5 mAac 066 166
Reserved 067 to 071 167 to 171
0/5 Adc 072 172
11.4
0/4 Adc 073 173
0/3 Adc 074 174
0/2 Adc 075 175
0/1 Adc 076 176
0/500 mAdc 077 177
0/300 mAdc 078 178
0/75 mAdc 079 179
0/50 mAdc 080 180
0/20 mAdc 081 181
0/10 mAdc 082 182
0/5 mAdc 083 183
Reserved 084 to 088 184 to 188
0/100 Hz
(Vac)
089 189
11.5
45/55 Hz
(Vac)
090 190
55/65 Hz
(Vac)
091 191
0/100 Hz
(Aac)
092 192
45/55 Hz
(Aac)
093 193
55/65 Hz
(Aac)
094 194
Reserved 091 to 099 191 to 199
(End of list) ‘---’ (see notes below)
(Custom selection) ‘uSEr’ (see notes below)
Select the desired code for your application, and check the following
sections for more information:
• for information on how to activate a code, see section 13.1
• to customize the input and output signals, see section 13.4
• to congure input for bipolar DC signals, see section 13.4
Notes
• Code ‘uSEr’ indicates that a user custom conguration is active, and it does not
match any of the listed codes This code is non-selectable, for information only.
Example: select code ‘011’ for 0/450 Vac=4/20 mA, the instrument reads code ‘011’.
Later, congure the input to 0/350 Vac=4/20 mA, this does not match a listed code,
and the instrument reads ‘uSEr’. Or change the output to 0/450 Vac=1/5 Vdc, this
does not match a listed code, and the instrument reads ‘uSEr’.
• Code ‘---’ identies the end of the list, it follows code ‘199’ and the list
continues with code ‘010’. Select ‘---’ to exit the list without applying changes.