PCE Instruments PCE Instruments | www.pce-instruments.com
9.  Access the ‘Conguration system
Key ‘SQ’ (<)
Conguration digits
Key ‘UP’ (5)
UP SQ
‘CONFIGURATION SYSTEM’
the ‘conguration digits’ indicate the active input signal range (see section 5)
keys ‘UP’ (5) and ‘SQ’ (<) to congure the instrument (see section 10.2)
UP SQ
Conguration digits
Opening the front cover reduces the security level. While front
cover is open :
the isolation between accessible parts and power supply is
reduced from ‘reinforced’ to ‘basic’.
• accessible parts are part of the input signal circuit.
• operations must be performed by qualied operators.
When correctly connected, the input signal terminal prevents the
front cover from opening.
ACCESS THE ‘CONFIGURATION SYSTEM’
Step numbers as referenced at the image below.
1. Remove the input signal terminal
make sure that there are no dangerous voltages at the output signal
terminals
2. Open the front cover
3. Congure the instrument
locate the ‘conguration digits
locate the ‘UP’ (5) and ‘SQ’ (<) keys
BLOCK ACCESS TO THE ‘CONFIGURATION SYSTEM’
4. Close the front cover
remove the input signal terminal (if it was placed)
5. Connect the input signal terminal
when connected at place, check that the front cover can not be
opened
if they were removed, connect the two output signal terminals
3
2
4