Program control function:
» Short press PROG to enter the program control mode, short press PROG again to enter the next program
control setting, a maximum of 6 groups of running programs can be set, that is, P1-P6 displays P1 by
default;
» Short press the left knob, set the time, rotate the left knob, can adjust the timing time. If the time is
00:00, the current program runs for a long time;
» Short press the right knob, set speed, rotate the right knob, can adjust the set speed;
» Short press SWITCH key to Switch the running direction of magneton in program-controlled mode;
» After completing the 6-section program control data setting, press the START/STOP button, and the
instrument will start to run according to the set program. Press again to stop running.
MAINTAINING
FAULT ANALYSIS
» Do not put electronic devices in detergent while cleaning.
» When cleaning the machine, do not handle hard objects to prevent damage to the surface of the
machine.
» Unplug the power cord if it is not used for a long time, and place the instrument in a cool and dry place,
and cover with dust bag.
» If the machine is not used for a long time, it should check whether all functions are normal before it is
restarted.
Fault Possible cause Solution
Unable to start normally The power cable is not tightly
connected
Check whether the power cable
is tightly connected or connected
to another power socket
Instrument does not stir/stir
when jumpers
» Motor fault
» The bottom of the beaker is
uneven and out of position
» Wrong speed adjustment step
» The actual voltage is within
the rated voltage range
» Check the motor wiring or
replace the motor
» Lay the bottom of the
container at, reposition the
beaker and start again
» Gradually adjust from low
speed to high speed
» Adjust voltage to rated
voltage range
There is no vortex in the liquid
after operation
» The rotational speed is
improperly set
» The viscosity of the stirred
mixture was too high
» Resetting speed
» Reduce the viscosity of the
mixture