19
SETUP
COMM PRO
SCPI or modbus communication setting
BUTTON VOICE
Enable or disable button sound
ALARM VOICE
Enable or disable alarm sound
BRIGHTNESS
Screen brightness adjustment from 10% to
100%, with a 10% adjustment each time.
BAUD
The baud rate setting range of the
communication device is from 115200 to 1200;
8 baud rates are available for selection.
RESET
Restore the settings on the panel to factory
defaults.
ADDRESS
The parallel machine address can be selected
from 1 to 10 to set the local machine's parallel
address.
Figure 4-7-1 system setup page
4.7.2 Parameter Setting
Select System Settings, then choose Parameter Settings. Press the Enter
key to enter the Parameter Settings options, as shown in the table below.
The parameter settings of the electronic load are as follows:
Menu
Settings
Description
Paramet
er setup
OVP
If the input voltage exceeds the set safety threshold
(OVP point), the OVP protection will be triggered.
OCP
If the input current exceeds the set safety threshold
(OCP point), the OCP protection will be triggered.
OPP
If the input power exceeds the set safety threshold
(OPP point), the OPP protection will be triggered.
RISE
SLOPE
The setting value for the rise rate of input voltage or
current during load operation.