Menu Operations
41
choose the "
REVOKE
" option to cancel the calibration.
Figure 5.2-28 Calibration Preview and Apply Prompt
◼
Method 2: Software Calibration Using Osee Calibrator
Calibration Preparation
:
Connect the USB interfaces of the control computer to both the target monitor
and the calibration probe through the calibration cable.
Connect the first USB interface of the control computer to the
CALIBRATE
interface of the OSEE monitor, and connect the second USB interface to the
calibration probe(the probe usually comes with a dedicated calibration cable),
as shown in Figure 5.2.28, and power on the devices.
Figure 5.2-29 Device Connection-Osee Calibrator Calibration
When using Osee Calibrator to calibrate the monitor, you can add a center marker
tool by selecting
ADD NEW TOOL
→
Frame
→
Center
, and activate it, then you