Microlife PF 200 BT handleiding

143 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 6 van 143
4
Press memory scroll buttons 6 to set the PEF value to be used as the reference
maximum of PEF traffic light.
Press time button to confirm selection and proceed to the next setting menu. Press O/I
button 1 to keep the current setting and exit settings menu.
Device date and time setup
The last setting in settings menu is device date and time, the date and time display
symbols will blink to allow adjustments, starting with year, then month, day, hour, and
minute.
Press Memory Scroll buttons 6 to adjust the value. Press TIME button AS to confirm
value and proceed to the next adjustment. Press O/I button 1 to keep the current setting
and exit settings menu.
5. Measurement preparation
Check the following items in preparation of measurement:
Take the measurement in positions of standing or sitting upright to obtain accurate
readings.
Check if the device has been cleaned or disinfected in case the device was previ-
ously used by another patient.
6. Measurement operation
Standalone measurement mode operation
1. Press O/I button 1 to turn the device on.
2. When device turns on, the previous measurement result AK will appear on the display
(0 if there is no data). The device will then sound two short beeps and show blinking
arrows AQ on display. The display will blink to indicate it is ready for measurement.
3. Follow these steps to perform a measurement correctly.
You can perform the measurement while standing or sitting upright. For better
comparison of your data, you should always perform the measurement in a
similar position.
Hold the device with both hands on the two textured compartments.
Cover the mouthpiece tightly with your lips. Ensure the lips form an airtight seal
around the mouthpiece 5.
NOTE: Do not bend your neck nor obstruct the mouthpiece with your tongue.
Inhale completely and hold your breath for a moment.
Blow into the measuring tube 4 as hard and as fast as you can.
NOTE: Avoid slow and sustained exhalation to obtain accurate reading.
Device will sound a long beep to indicate the exhalation is measured; the PEF
value is displayed for 3 seconds followed by FEV1 value
AO
. Then, the device will
sound two short beeps to indicate that it is ready for a new measurement.
Repeat measurement as needed. It is recommended to perform three or more
measurements sequentially.
4. After each measurement, the actual measurement result is shown followed by the
highest reading of your current measurement session.
5. Press the O/I button 1 to finish the measurement session and record the highest
measurement value in the device memory. The device will display the saved PEF
and FEV1 values with the memory slot number AN.
6. After saving the measurement result, the device will activate Bluetooth® datalink if
automatic Bluetooth® activation is set to «On». To Turn device off without Blue-
tooth
®
data transfer, press O/I button 1 again.
7. Clean the measuring tube and mouthpiece after use.
7. Measurement interpretation
Self-assessment using the PEF traffic light indicator
The colored bars on the left of the display is the PEF Traffic Light Indicator 9, which
provides a quick visual reference comparing your PEF result versus a reference baseline
value. As your PEF result raises or lowers, the arrow on the display will show the
measurement as either within the normal (green), borderline (yellow) or danger (red)
range.
Green Zone OK: If your measurement is in the green zone, your condition appears
to be under control. Continue your treatment plan.
Yellow Zone – Caution: If your measurement is in the yellow zone, measure more
frequently and follow your treatment plan.
Red Zone – Alert: If your measurement is in the red zone, your condition is serious.
Act as discussed with your physician or seek emergency medical treatment.
Baseline reference of the PEF traffic light indicator
There are two different settings of baseline reference value available for the PEF traffic
light indicator:
Memory: Using your personal highest PEF result in the device memory as the base-
line for the traffic light indicator zones.
Customized: Set a customized PEF baseline (e.g., predicted reference) value
manually on the device or use the compatible software for the traffic ligh indicator
zones.
The green zone ranges between baseline or predicted value (MAX) and 80% of that
reading. To determine the yellow/green border zone, multiply the baseline MAX by
0.8 [Example: 500 l/min x 0.8 = 400 l/min].
The yellow zone ranges between baseline or predicted value (MAX) and 60% of that
reading. To determine the yellow/red border zone, multiply the Baseline MAX by 0.6.
[Example: 500 l/min x 0.6 = 300 l/min].
The red zone ranges below 50% of baseline or predicted value.
8. Memory function operation
During a measurement session, the device will measure multiple exhalation readings
and automatically record the highest reading when the session ends (for example: 1
session is 1 on–off cycle). The device memory stores 400 readings.
Review measurement readings in memory
To review the measurement readings in device memory, switch the device on and press
Memory Scroll buttons 6. Press «<» of the Memory Scroll button to review the second
latest readings; press «>» works in the opposite direction.
Memory capacity low
When the number of readings stored in the device memory reaches 390 sets or more,
the device will display AN blinking after switched on to indicate that the device memory
capacity is low.
Memory full
When the number of readings stored in the device memory reaches 400 sets, the device
will beep and display AN «400» blinking when switched on to indicate that the device
memory is full. From this point on, any new measurement readings will replace and over-
write the oldest reading in the device memory.
Delete the most recent reading in memory
To delete the most recent data in the device memory, while the device is in memory func-
tion, press Memory Scroll (both «<» and «>») buttons 6 simultaneously for 5 seconds
until the display shows «CLr» blinking. Press the Memory Scroll buttons 6 again to
delete the most recent reading in device memory. Wait until the «CLr» message on the
display elapses after 3 seconds to exit without change.
Delete all readings in memory
To delete all data in the device memory, while the device is in memory function, press
Memory Scroll (both “<” and “>”) buttons 6 simultaneously for 5 seconds until the
display shows «CLr» blinking. Press O/I button 1 to delete all readings in the device
memory. Wait until the «CLr» message on the display elapses after 3 seconds exit
without change.
Device date and time can also be set using companion computer or mobile soft-
ware.
Adult supervision is required for taking measurements of children, elderly
person, and patients with disabilities.
PEF traffic light indicator is not active when memory contains no PEF values.
This device is preset with a customized baseline of 500 L/min.
Setting a predicted reference PEF value, creating a treatment plan and modi-
fying your traffic light zone settings should only be done under direction of a
physician (or other licensed health care practitioner). If you would like to use this
feature, determine your customized zones with your doctor and record them on
the traffic light card (enclosed).