Kimo C 310 handleiding
Handleiding
Je bekijkt pagina 8 van 34

Type Size Description Format
Date 4 bytes Year (2 bytes)
Month (1 byte)
Day (1 byte)
Byte 2
(month)
Byte 1
(day)
Byte 4
(year
most-
signicant
byte)
Byte 3
(year least-
signicant
byte)
Example with 31/03/2014 (0x07DE031F) 0x03 0x1F 0x07 0xDE
Type Size Description Format
Hours 5 bytes Hour (1 byte, digital)
Minute (1 byte, digital)
Second (1 byte, digital)
Unused (1 byte, indifferent)
Byte 2
(minute)
Byte 1
(hour)
Byte 4
(unused)
Byte 3
(second)
Example with 22h 35 min 06sec (0xXX062316) 0x23 0x16 0xXX 0x06
Type Size Description Example
Serial number 8 bytes Class (1 byte)
Range (1 byte)
Year (2 bytes)
Month (1 byte)
Number (3 bytes)
'3' (0x33)
'F' (0x46)
13 (0x000D)
8 (0x08)
98765 (0x0181CD)
Format
Byte 2 (range) Byte1 (class) Byte 4 (year) Byte 3
(year)
Byte 6
(number)
Byte 5
(month)
Byte 8
(number)
Byte 7
(number)
0x46 0x33 0x00 0x0D 0xCD 0x08 0x01 0x81
Example with 3F13898765: 0x0181CD08000D4633
Alarms and relays status – Modbus code : 7000
Encoded on 4 bytes (U32)
Byte 2 Byte 1
b8 – b15 b7 – b4 b3 b2 b1 b0
Unused Unused
Channel 4 Channel 3 Channel 2 Channel 1
Alarm state
*
(*)
1: the channel is in alarm state / 0: the channel is not in alarm state
Byte 4 Byte 3
b31 – b28 b27 b26 b25 b24 b23 – b20 b19 b18 b17 b16
Unused Relay 4
**
Relay 3
**
Relay 2
**
Relay 1
**
Unused Alarm 4
***
Alarm 3
***
Alarm 2
***
Alarm 1
***
(**)
1: the relay is triggered / 0: the relay is not triggered
(***)
1: the alarm is activated / 0: the alarm is deactivated
8
Modbus
Bekijk gratis de handleiding van Kimo C 310, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.
Productinformatie
| Merk | Kimo |
| Model | C 310 |
| Categorie | Niet gecategoriseerd |
| Taal | Nederlands |
| Grootte | 3908 MB |