Pro Series Ozone Generator
5.2 Replace Ozone Modules
The CD40, CD70, and CD100 have 3, 5, and 7 ozone modules,
respectively. Replace each ozone module whenever an indicator
light goes out, or after 15,000 hours of operation, whichever comes
first.
If green indicator light(s) are lit but dimly glowing, the ozone module
may be producing less ozone due to contamination within the plasma
gap ozone chamber. Replace the module even if the green light is
still on.
5.3 Verify Ozone Flow
Periodically check the flow meter for proper flow. Always remove the
flow meter after confirming proper flow. Always recheck the flow rate
if replacing any components in the flow line.
5.4 Replace the Ozone Tube
WARNING
Do NOT touch the ends of the Ozone Tube when replacing. Trace
amounts of nitric acid may be present and could prove harmful if
touched or ingested.
Replace the ozone tube annually, or as needed. Inspect tube
regularly for cracks or wear and replace as necessary.
If there is evidence of water leaking past the gas line check valve
(on the generator side), shut off the pump immediately. Replace the
ozone tube and the gas line check valve.
If water entered the generator, allow the unit to dry completely before
restarting.
NOTE: Evidence of water in the ozone generator may void the warranty.