1) With the controller running, press and hold the
and
buttons simultaneously for 2 seconds until the SV screen dis-
plays and the CO2 screen displays 400ppm (default), then
release both buttons.
2) Then, press or to adjust the parameter.
3) After confirmation, press and hold the button for 2 sec-
onds to start calibration, the CO2 screen will display and flash.
4) Calibration takes 1 to 20 minutes. Please be patient.
5) It is recommended to calibrate to 400ppm in a ventilated out-
door environment or to a number that matches the known CO2
rated value.
6) Calibration Result:
If the CO2 screen displays and the SV screen displays
, the calibration is successful. Press any button to exit.
If the CO2 screen displays and the SV screen displays
, the calibration has failed. Perform the calibration again
or press any button to exit.
13
NOTE: Please place the CO2 sensor in an environment with a
stable CO2 level, otherwise the calibration time may be prolonged
or the calibration may fail. If the CO2 level in the current environ-
ment fluctuates greatly, please use the general calibration.
NOTE: When the CO2 sensor is performing automatic calibration,
with the SV screen displaying and flashing, do not exhale
into the sensor.
3.5.2 Professional Calibration
PPM
%RH
PPM
%RH
PPM
%RH
PPM
%RH
%RH
PPM
PPM
%RH
3.6 Alarm Setting
3.6.1 Enable the audible alarm
1) When running, press and hold the button for 2 seconds to
enter the setting mode.