Only for CO2 devices, such as a CO2 regulator or generator. (For
Human Mode, see section 3.4.3)
NOTE: Please note that the photo sensor only works for the CO2
outlet. When there is light, it will turn the CO2 outlet on/off based
on the set values; when there is no light, it will keep the CO2 outlet
off regardless of the set values.
3.4.1 Plant Mode without Fan
07
Press and hold the button to
enter the setting mode.
(CO2)
(CF)
Press or to select CO2, which means that only the CO2
outlet is enabled, then press the button.
(1200ppm)
(Con)
Press or to adjust the CO2 start value, then press the
button.
(1600ppm)
(CoFF)
Press or to adjust the CO2 stop value, then press the
button. Please note that the CO2 stop value should be greater than
or equal to the CO2 start value. (CO2 Stop value ≥ CO2 Start value)
(0ppm)
(CAL)
CAL=calibration; Calibrate the value only if it is confirmed to
be biased. The default value is .