IFM SUH251 handleiding
Handleiding
Je bekijkt pagina 32 van 45

SUHxxx Ultrasonic flow meter without display
32 © ifm electronic gmbh 11661449 / 01 07 / 2025
10.3.5 Low flow cut-off
Low flow quantities can be ignored using the parameter LFC (Low flow cut-off). Flow below the LFC value
is evaluated by the sensor as standstill (Q = 0).
The LFC value influences:
▪ The switching signal for flow
▪ the analogue signal for flow
▪ The frequency signal for flow
▪ the consumed quantity monitoring (switching or pulse signal for flow)
▪ The totalisation of the consumed quantity by the totaliser.
▪ the memory values for minimum and maximum flow
+ LFC
+ Q
t
0
- Q
- LFC
1
Fig.14: Low flow cut-off
+LFC: Minimum flow in positive flow direction
-LFC: Minimum flow in negative flow direction
1: Flow which is evaluated as standstill
10.3.6 Medium
The sensor provides various characteristic curves for the respective media. They can be selected via the
MEdI parameter.
Selectable values:
▪ H2O: Water
▪ OIL46: High-viscosity oils (viscosity: 30…68 mm²/s at 40°C / 30…68 cSt at 104°F)
10.3.7 Flow direction
The positive flow direction can be defined by the user. This setting affects the following functions:
Consumed quantity monitoring (totaliser function) [ 24]
Switching signal for flow direction [ 23]
Analogue signal [ 26]
An arrow with the text “flow direction” on the device indicates the positive flow direction (factory setting).
The direction of the flow rate measurement can be reversed using the parameter Fdir:
Fdir Direction of flow
+ Flow direction in case of factory setting
- Flow direction contrary to the factory setting
Bekijk gratis de handleiding van IFM SUH251, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.
Productinformatie
Merk | IFM |
Model | SUH251 |
Categorie | Niet gecategoriseerd |
Taal | Nederlands |
Grootte | 3640 MB |