SU2x5x SU6x5x SU7x5x SU8x5x SU9x5x Ultrasonic flow meter
10
u Tighten the two adapters in opposite direction using a torque wrench applying a tightening torque
of 30Nm. Fix the device at the hexagon using a spanner.
For SUx6x1: Use the hexagons on both sides of the device for fixing.
ATTENTION
Sensor damage
w Higher tightening torques (>50Nm) will permanently deform and damage the device.
u To tighten the adapters on the sensor, use a screwing tool that allows you to set a fixed
tightening torque.
5.2 Interference
Structures in the pipe, bends, valves, reducing pieces and the like affect the function of the unit.
u Adhere to the distances between sensor and interference.
S S
5 x D
1 x D
F
Fig.2: inlet and outlet sections
D: outside diameter of the pipe
S: interference
5.3 Installation position
5.3.1 Recommended installation position
u Install the unit so that the measuring pipe is always completely filled.
u Install in front of or in a rising pipe.
If air bubbles can form in the pipe system:
u In case of horizontal installation, mount the sensor with the top of the device to the side of
the pipe (A).