10 Commissioning
After power on and expiry of the power-on delay time, the unit is in the normal operating mode. It carries
out its measurement and evaluation functions and generates output signals according to the set parame-
ters.
During the power-on delay time, the outputs are in the following status according to the set parameters:
Switching output:
• Active (“high” with normally open function; “low” with normally closed function)
Analogue output:
• 0 mA with current output (I or I
neg
)
When an IO-Link master is connected, the device automatically goes from SIO mode (standard
input-output) into IO-Link mode if the port of the master is set to IO-Link mode.