PG17xx Electronic manometer
22 © ifm electronic gmbh 11601190 / 01 07 / 2025
8 Operating and display elements
4
2
0
6
8
10
psi
psi
Output
Status
TEMPbar
60
40
20
0
80
100
120
140
bar
Enter
-1
1
2
3
4
8
6
7
5
Fig.9: Operating and display elements (example device type with a measuring range of -1...10 bar; display layout L1)
1: Analogue display of the current system pressure in bar (outer scale) and psi (inner scale)
2: LED ring for colour visualisation of the analogue display.
• The range between 0 and the measuring range end value is divided into 72 LED segments.
• A white LED segment extending the pointer indicates the current process value.
• 3 colour ranges can be configured
48
. Example:
– Upper colour range: 3.6...6 bar, green
– Middle colour range: 2.6...7 bar, yellow
– Lower colour range: -1...8.8 bar, red
• The display layout of the LED ring can be adjusted
47
. Depending on the setting: Lag indicator function for
minimum and maximum pressure values.
3: LED pointer: White LED, not visible when the unit is not powered. Indicates the current pressure value between 0
and the final value of the measuring range in 72 LED steps. The LED segment next to the pointer (pointer exten-
sion) lights white.
4: Switching status LED for OUT1: lights yellow if output 1 is active.
5: Digital display (4-digital alphanumeric display):
• Indicates the current value. Depending on selection under SELd: pressure or temperature.
• Indication of the parameters and parameter values.
6: LED provides complementary information about the digital display:
• If SELd = PRES, bar (mbar) or psi lights, depending on what is set under uni.P.
• If SELd = TEMP, TEMP lights. The unit set under uni.T is displayed.
7: Status LED The LED lights depending on the device status:
• green in normal operation
• red in the event of a warning
57
or error
58
8: Optical operating keys (device keys) for changing the display and for parameter setting
32
. The device keys
can be locked
45
.
The term “device keys” refers to the touch buttons Enter,
▲
and
▼
.
The touch button Enter is indicated by the
●
icon below.
The sensor fields of these optical keys are operated simply by touching the viewing glass. The
touch button must be completely covered to be activated. Slow covering (e.g. liquid flows over
the display) does not activate the touch button.