Electronic manometer PG17xx
49
Process values for temperature can only be simulated via the IO-Link interface.
The simulation time is adjustable: 1, 2, 3, 4, 5, 10, 15, 20, 30, 45, 60 minutes.
During the simulation:
The simulated value or the simulated event (e.g. UL) appears in the digital display, alternating with
“SIM”.
The LED pointer points to the simulated value in the LED ring or behaves according to the
simulated event (e.g. LED pointer off in case of an error).
The switching status LED lights according to the simulated values.
The status LED continues to signal warning and error states of the current application.
The digital display does not show warning or error messages of the current application.
The calibration values [coF.P] and [CGA.P] are set to zero. After the simulation, the device reverts
to the set values.
The configured damping values [dAP.P] and [dAA] are used for the simulation.
If the simulation is started via IO-Link, it can also only be finished via IO-Link. If you try to stop
the simulation using the device keys, [C.Loc] will be displayed.
10.7.3.1 Parameterisation via unit keys: Simulation
u Go to the [EF] > [SIM] menu.
u Select [S.PRS] and set the pressure value to be simulated or a diagnostic case (e.g. OL).
u Select [S.diA] and set the diagnostic case to be simulated.
Diagnostic cases that can be set: Main menu and submenus (Ò/27).
u Select [S.Tim] and set the time of the simulation in minutes.
u Select [S.On] and set the desired function:
[On]: The simulation starts. The values are simulated for the time set under [S.Tim]. Cancel the
simulation using the [●] key.
[OFF]: The simulation is not active.
10.7.4 Binary data transmission (BLOB)
The device offers a function for reading binary data from the device as one large file (BLOB = Binary
Large Object).
The data is exported as a BIN file.
This requires a software tool (e.g. ifm moneo) that supports the IO-Link BLOB interface.
The BIN file contains the following logbook information:
device information for identification
number of operating hours
event logging:
event history with time stamp
event code
event description
event frequency
device restarts
Only the last 50 events are saved.