OPERATION
47
Clear
OVP/OCP
protection
The OVP and OCP protection can be
cleared after it has been tripped by
clicking Shift key + ALM CLR key.
+
Set to C.V. Priority Mode
When setting the power supply to constant voltage mode, a current
limit must also be set to determine the crossover point. When the
current exceeds the crossover point, the mode switches to C.C. mode.
For details about C.V. operation, see page20.
C.C. and C.V. mode have two selectable slew rates: High Speed
Priority and Slew Rate Priority. High Speed Priority will use the
fastest slew rate for the instrument while Slew Rate Priority will use
a user-configured slew rate.
Background
Before setting the power supply to C.V. mode, ensure:
The output is off.
The load is connected.
Steps
1. Press the Menu key followed by clicking
on Output to enter the Output page.