Parallel/Series Operation
127
Series
Configurati
on
Before using the power supplies in series, the master
and slave units need to be configured.
1. Configure the OVP and OCP settings for the master
unit.
2. Enter the Power ON Configuration settings and set
the Tracking configuration for the master and each
connected slave.
3. If using voltage sense, disable local sense in the
Power ON Configuration settings.
4. Cycle the power on the units (reset the power).
Note
Set Tracking to Local to return the slave units to local
(independent) operation.
Master-
Slave
Operation
Only operate the power supplies in series if the units
are configured correctly.