GPP-1323/GPP-1205 User Manual
126
Series
Sense
Connection
For remote sense connections, connect the sense
terminals as shown below:
a. Connect the Master S+ terminal to the positive
potential of the load.
b. Connect the Master S- terminal to the positive output
terminal of the slave unit.
c. Connect the slave S+ terminal to the positive terminal
of the slave unit.
d. Connect the slave S- terminal to negative terminal of
the load.
Steps
1. Ensure the power is off on both power supplies.
2. Connect the master and slave unit in series as shown
above to either increase the voltage output or to create a
positive and negative output. Remember that how the
units are grounded depends on the configuration of the
series connection.
Note
Ensure load cables have sufficient current capacity.
Series Operation