Repair
F1, F2 and F3 (3-phase mains supply power sys-
tem N2)
Required materials:
A slotted screwdriver for opening the
fuse compartment
Replacement fuses in accordance with
the details specified in the Technical Data
on page 9.
Open the respective fuse compartment.
Insert the slotted screwdriver into the slot
on the fuse compartment and turn it
counterclockwise.
Pull out the fuse cover as soon as it pro-
trudes slightly from the housing.
Remove the defective fuse from the fuse
cover.
Insert the new fuse into the fuse cover.
Insert the fuse cover with the fuse back
into the fuse compartment. The two tabs
on the fuse cover must be inserted into
the recesses in the housing.
Insert the slotted screwdriver into the slot
on the fuse compartment and turn it
clockwise until the fuse compartment is
closed.
F
LN
(mains power supply system single-phase N1)
You require replacement fuses in accor-
dance with the details specified in the Tech-
nical Data on page 9.
Open the fuse compartment. Press to-
gether the plastic wings of the fuse link
to this end and pull it up slightly.
As soon as the fuse link protrudes a little
bit, pull it out.
Remove the defective fuses from the
fuse link.
Insert the new fuses into the fuse link.
Insert the fuse link with the fuses into the
fuse compartment. The tab on the fuse
cover must be inserted into the recesses
in the housing.
Press down the fuse link until it snaps
into place.
12 Repair
If your instrument requires repair, please
contact our service department, see “Con-
tact, Support and Service” on page 20.
Note
Loss of warranty and guarantee
claims!
Unauthorized modification of the tes-
ter is prohibited. This also includes
opening the meter.
If it can be ascertained that the tester
has been opened by unauthorized
personnel, no guarantee claims can
be honored by the manufacturer with
regard to personal safety, measuring
accuracy, compliance with applicable
safety measures or any consequen-
tial damages.
If the guarantee seal is damaged or
removed, all guarantee claims are
rendered null and void.
The device may only be repaired or opened
by authorized, qualified personnel who are
familiar with the associated dangers.
Original replacement parts may only be
installed by authorized, qualified personnel.
The instrument may not be placed back into
operation until troubleshooting and repair
have been performed, and calibration and
dielectric strength have been tested and
approved at our factory or by an authorized
service center.