LABKON Series
_________________________________________________________________________________________________
26 GMC-I Messtechnik GmbH
3.12.4 4.5 V Calibration
Press “Enter” key to verify and then the display will show as follows:
Enter the value you read from the multimeter, which is retained to four decimal places. If necessary, press “Clear” key to
remove the wrong input value of current digit. Then press “Enter” key to confirm and exit to calibration menu. Now, voltage
calibration is completed.
3.13 CC Mode Calibration
In this mode, three current points: 0.5 A, 6 A, 9 A (take LABKON P500 35 V/14.5 A, K148A as an example) should be
calibrated. A complete overview of calibration checkpoints for all LABKON models is listed at the end of this manual!
3.13.1 Wiring
3.13.2 0.5 A Calibration
Press “Enter” key, you will see the following figure in the display:
Enter the value you read from the multimeter, which is retained to four decimal places. If necessary, press the “Clear” key to
remove the wrong input value of present digit. Then press “Enter” key to confirm.