4.02.4 E.A.C. Collector Cell / HEPA Filter Removal
Hood Operation Free Standing Ventless Hood
FSH-2 & FSH-2A-99
4.02.5 E.A.C. Collector Cell / HEPA Filter Installation
E
A
C
1
E
A
C
2
Unlatch access cover & lift off.
Grasp E.A.C. Collector Cell / HEPA Filter & pull straight out, on a slight upward angle.
Turn OFF hood power ...
E
A
C
Pin must align & fit
thru hole in rear of
cell.
Airflow
arrow
must point
up
2
1
Ensure the airflow indicator arrow
points [UP] & the cell’s contact pad
faces to the right side.
Align cell (or filter) in guides and
slide into hood. If installed properly,
cell is flush with hood front. If not,
cell is not installed incorrectly.
Replace access cover and latch.
E
A
C
NOTE: HEPAFilter models ONLY charcoal filter will be located below the HEPA filter, so that air passes through it first.
When installed, the filter must properly engage a proofing switch located on the back wall.