Acoustic Imager
Menus
c. Electrical
• Discharge Type: pre-dened partial discharge type by
on-device data analysis and highest likelihood. Option
for end-user to modify.
• Voltage / Frequency: Open numeric entry.
• Ambient Temp: Open numeric entry.
• Relative Humidity (%): Open numeric entry.
• Weather Notes: Open text eld.
d. Mechanical (ii915)
• Component Type: Pre-dened list of mechanical
components.
• Speed (RPM): Open numeric entry.
• Severity: High, Medium, Low (Default:
Undeterminded).
For QR code-based Asset ID recognition:
4. Tap to use the camera to capture and read QR code.
Inspection Status: Asset Inspection Status helps you to select
and assign the status of inspection:
• Action required: Yes, No (Default: Undetermined).
• Action priority: High, Medium, Low (Default:
Undetermined).
• Action notes: Open text eld.
5. Tap
to close the Tag Note menu.
Acoustics
The Acoustics menu shows all the available settings for adjustment.
Show PD Count: On or O
The PD Count (Partial discharge pulse count) value is an indication
of the number of pulses/minute that the Imager detects.
Show dB Scale: On or O
You can choose to show or hide the dB scale. Turn o the dB scale
to see a larger visual area on the display.
Min / Max dB
The minimum/maximum decibel (dB) settings determine the sound
level (intensity) that shows on the SoundMap
TM
. Decibel level
thresholds help you to visualize leaks in challenging conditions, for
example, very small leaks or a lot of background noise in the same
frequency range as a leak. For more information, see Proles.
Auto: Automatically adjusts the Color Palette Scale to the minimum/
maximum decibel value for the received sound pressure.
Manual: The Color Palette Scale is a user-dened minimum/
maximum decibel value. Levels above the maximum value show
on the display with the same color as the maximum value. Levels
below the minimum value do not show on the display.
When Manual is selected, use the +/- to adjust. Or, tap on the dB
SPL scale and use the slider to manually adjust the minimum and
maximum in decibel values of the Color Palette Scale.
High Frequency (ii915)
Toggle between the standard frequency range up to 52 kHz and the
expanded frequency range up to 100 kHz.