DWL-1300XY & DWL-1500XY Instruction Manual
5
2.5. Calibration Mode
1. Turn Off device. Next, Press & Hold “ON/OFF” button & then “ALTERNATE ZERO”
button simultaneously, release both buttons when displays ashing “01”.
2. Place the device to position 1. Press
“MODE” button once and wait until count-
down reaches “0”, LCD screen displays ashing “02”.
3. Turn the device 180° (reverse position) to position 2. Press
“MODE” button once
again and wait until countdown reaches “0”. LCD screen automatically switch to
measuring mode once calibration step 2 is completed.
Figure 6. Calibration mode display in LED screen (a) Calibration Step 1, (b) Calibration Step 2, (c) Calibration countdown
4. Verify your calibration result by comparing the measurement value of current posi-
tion with the device after turning 180° (in reverse position). If displayed measurement
value differs > 0.001°, the device has not been calibrated properly (e.g. surface un-
evenness due to rough surfaces or dirt, etc.). Please put a device on smooth surfaces
to re-calibrate again.
Figure 7. Position 1
Figure 8. Position 2
a b c
Calibration tutorial video available on Digipas website, please visit the following link:
https://www.digipas.com/support/video.php