3.2. OPERATION:
3.2.3. M
measurement:
M
measurement at 250V
DC
:
The measurement is made at a voltage of 250V
DC
above 1M measured. Short-
circuit current 5mA
DC
.
There is a change of reading coefficient for rising values: 50M (changes to X10
reading), and a change of reading coefficient for falling values: 11M (changes to
X1 reading).
M
measurement at 500V
DC
:
The measurement is made at a voltage of 500V
DC
above 1M measured. Short-
circuit current 5mA
DC
.
There is a change of reading coefficient for rising values: 50M (changes to X10
reading), and a change of reading coefficient for falling values: 11M (changes to
X1 reading).
M
measurement at 1000V
DC
:
The measurement is made at a voltage of 1000V
DC
above 10M measured.
Short-circuit current 5mA
DC
.
There is a change of reading coefficient for rising values: 500M (changes to X10
reading), and a change of reading coefficient for falling values: 110M (changes
to X1 reading).