AWT35-100109 Rev AB 23/29
OV.IZSM
DSP.OVR
DSP.OVR
Sets the weight type as current initial zero
point when current weight is outside the
IZSM range:
DSP.OVR = if the initial zero is over range,
display the 0
-----
error;
WEIGHT = current weight;
CAL.ZRO = calibration zero;
LAST.Z.T = last zero/tare value.
WEIGHT
CAL.ZRO
LAST.Z.T
SAZSM 0~100 20
Sets the ZERO key range:
0 = no limit;
1~100 = (initial zero point) ±1%FS ~ (initial
zero point) ±100%FS
ZRO.PNT AZSM 0~100 56
Sets the Automatic Zero Tracking weight
range:
0 = 0 div. Disabled;
1~100 = ±(0.2+0.05*(1~100)) div/sec
FILTER
L1/ L2/ L3
L3
Filtering settings
FUNC
HOLD
YES / NO
YES
Yes = enables the Hold function;
COUNT
YES / NO
YES
Yes = enables the Count mode;
PERCNT YES / NO YES
Yes = select the displayed % value
format 100% , 100.0% or 100.00%
BMI YES / NO NO
No = disabled.
See MS 1000-LCD Operation Manual.
COMPAR
YES / NO
YES
Yes = enables the Compar mode;
ACCUMU YES / NO YES
Yes = select the Accumulation mode:
MANUAL: add up the current value to
memory when the TOTAL key is pressed.
AUTO: add up the current value to
memory when the scale is stable and
weight is over (NLD.RNG
)
GEO.CAL
YES YES
Yes = enables the Geographical
Adjustment Factor in the Calibration
Menu.