Barco MUIP-2213 handleiding

38 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 28 van 38
R5916765 /01 MUIP-221328
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment guidance
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m 30 A/m Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
Conducted RF
IEC 61000-4-6
3 Vrms (6 Vrms in ISM
bands)
150 kHz to 80 MHz
3 Vrms (6 Vrms in ISM
bands)
If there is a mark on the
device as shown on the
right , interference may
occur.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
IEC 61000-4-39 8 A/m @ 30kHz (CW)
65 A/m @ 134.2 kHz (2.1
kHz Pulse modulation)
7.5 A/m @ 13.56 MHz (50
kHz Pulse modulation)
8 A/m @ 30kHz (CW)
65 A/m @ 134.2 kHz (2.1
kHz Pulse modulation)
7.5 A/m @ 13.56 MHz (50
kHz Pulse modulation)
-
Immunity to RF wireless communications equipment
Test
frequency
(MHz)
Band (MHz)
Service Modulation
Maximum
power (W)
Distance
(m)
Immunity
test level (V/
m)
385 380 390 TETRA 400 Pulse
modulation
18 Hz
1.8 0.3 27
450 430 470 GMRS 460,
FRS 460
FM ± 5 kHz
deviation
1 kHz sine
2 0.3 28
710 704 787 LTE Band
13, 17
Pulse
modulation
217 Hz
0.2 0.3 9
745
780
810 800 960 GSM 800/
900, TETRA
800, iDEN
820, CDMA
850, LTE
Band 5
Pulse
modulation
18 Hz
2 0.3 28
870
930
1720 1700 1990 GSM 1800,
CDMA 1900,
GSM 1900,
DECT, LTE
Band 1/3/4/
25, UMTS
Pulse
modulation
217 Hz
2 0.3 28
1845
1970
2450 2400 2570 Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
217 Hz
2 0.3 28
5240 5100 5800 W LAN
802.11 a/n
Pulse
modulation
217 Hz
0.2 0.3 9
5500
5785
Important information