27R5916765 /01 MUIP-2213
Emissions test
Compliance
Electromagnetic environment
Guidance
directly connected to the public low-
voltage power supply network that
supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class D
1
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies
This MUIP-2213 complies with appropriate medical EMC standards on emissions to, and interference from
surrounding equipment. Operation is subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference received, including interference that
may cause undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna or equipment.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The MUIP-2213 is tested to be used in the electromagnetic environment specified below. The customer or the
user of the MUIP-2213 should assure that it is not used in an environment that exceeds the listed test levels
and limits.
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC 61000-4-5
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at
70% residual voltage for
25 periods at
Voltage interruptions: 0%
residual voltage for 250
periods at
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at
70% residual voltage for
25 periods at
Voltage interruptions: 0%
residual voltage for 250
periods at
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the MUIP-2213 requires
continued operation
during power mains
interruptions, it is
recommended that the
MUIP-2213 be powered
from an uninterruptible
power supply or a battery
Important information
1. Active power for MUIP-2213 is less than 75 W