4.1 Safety information
General recommendations
Read the safety and operating instructions before operating the device.
Retain safety and operating instructions for future reference.
Adhere to all warnings on the device and in the operating instructions manual.
Follow all instructions for operation and use.
Electrical Shock or Fire Hazard
To prevent electric shock or fire hazard, do not remove cover.
No serviceable parts inside. Refer servicing to qualified personnel.
Do not expose this apparatus to rain or moisture.
Modifications to the unit
Do not modify this equipment without authorization of the manufacturer.
Preventive maintenance
Periodic maintenance inspections are essential to keep the LCD User Interface in optimum condition and
ensure safe operation.
With the LCD User Interface disconnected from the mains, perform the following periodic check:
Check the integrity of the power cord and inspect its routing, so that it is not under the risk of being
punched or cut.
General recommendations:
Keep the LCD User Interface clean to prolong its operational lifetime.
LCD panel performance may deteriorate in the long term.
Periodically check that it is correctly operating.
Periodically check the tightness of the VESA mount screws. If not sufficiently tight, the LCD User Interface
may detach from the arm, which may result in injury or equipment damage.
Type of protection (electrical):
LCD User Interface provided with external power supply: Class II equipment.
LCD User Interface provided without external power supply: Class I or II, depending from the Class of the
external power supply used.
Degree of safety (flammable anesthetic mixture):
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
The equipment is not operable if the air oxygen content is above 25%.
Non-patient care equipment
Equipment primarily for use in a health care facility, including the patient area.
The equipment should not be used with life support equipment.
The user should not touch the signal input ports (SIP) / signal output ports (SOP) and the patient at the
same time. The rear cover must be kept closed and screwed.
Child safety
Equipment not suitable for use in locations where children are likely to be present.