5 ADAM-5000 Series User Manual
Chapter 1 Overview
Technical Specifications of ADAM-5013
ADAM-5013 RTD Input Resistance Calibration
1. Apply power to the module and let it warm up for about 30 minutes.
2. Make sure that the module is correctly installed and is properly configured for
the input range you want to calibrate. You can use the ADAM utility software to
assist in this.
3. Connect the correct reference self resistance between the screw terminals of the
ADAM-5013 as shown in the following wiring diagram. Table 2 below shows the
correct values of the span and zero calibration resistances to be connected. Ref-
erence resistances used can be from a precision resistance decade box or from
discrete resistors with the values 60, 140, 200 and 440 ohms.
Figure 1.3 Applying Calibration Resistance
Table 1.3: Technical Specifications of ADAM-5013
Analog input channels 3
Input type Pt or Ni RTD
RTD type and temperature range
Pt -100 to 100°C a=0.00385
Pt 0 to 100°C a=0.00385
Pt 0 to 200°C a=0.00385
Pt 0 to 600°C a=0.00385
Pt -100 to 100°C a=0.00392
Pt 0 to 100°C a=0.00392
Pt 0 to 200°C a=0.00392
Pt 0 to 600°C a=0.00392
Ni -80 to 100°C
Ni 0 to 100°C
Isolation voltage 3000 V
DC
Sampling rate 10 samples/sec (total)
Input impedance
2 M
Bandwidth
13.1 Hz @ 50 Hz
15.72 Hz @ 60 Hz
Input connections 2, 3 or 4 wire
Accuracy ± 0.1% or better
Zero drift ± 0.015°C/°C
Span drift ± 0.01°C/°C
CMR@50/60 Hz 150 dB
NMR@50/60 Hz 100 dB
Power consumption 1.2 W