What to do if ... IDS WA System
20 ba76199e03 08/2018
7.2 Display of when a sensor is connected
7.3 The battery of the IDS WA-S adapter cannot be charged
– IDS WA-S adapter
or
IDS WA-M adapter
not identified
– Disconnect and reconnect the IDS WA-S adapter to
the sensor.
– Carry out a reset of the IDS WA-S adapter con-
nected to the sensor: Press the function key of the
adapter for more than 8 seconds
– Carry out a reset of the IDS WA-M adapter con-
nected to the meter: Disconnect and reconnect the
adapter to the meter.
– Carry out a reset of the meter (see operating man-
ual of your meter)
– Strong damping of the
radio signal
– Remove any obstacles in the radio link (e.g. doors,
panes etc.)
– Run the meter and sensor at a smaller distance
– Check the signal strength (in the menu Sensors
<F1__>
Cause Remedy
– The maximum number
of sensors or adapters
is exceeded
– Disconnect the IDS sensor
– Disconnect the IDS WA-M adapter
Cause Remedy
– Insufficient contact of the
IDS WA-S adapter in the
charging station
– Clean the contacts
– Fix the adapter with the retaining clip
– Battery exhaustively dis-
charged
– Carry out a reset of the IDS WA-S adapter in the
charging station (IDS WA-S Charger):
Press the function key of the adapter for at least 8
seconds
Cause Remedy