6-26
The default output is Active High (+5V). Touch the Feed Pump Line [High] button to change it to [Low] if your alarm/pump
requires an Active Low signal (0V) to turn on.
Touch [Save] to save changes and return to the Edit Chamber 1 screen.
Touch [Save] to save changes and return to the Edit Configuration screen.
Touch [Save] to save changes and return to the Select Configuration screen.
Touch [Select] to choose the selected Configuration and return to the Monitor screen.
Print Monitor Configuration
Touch the Print Monitor Configuration button to print the current monitor configuration. The configuration can then be
reviewed on the YSI 2901 Printer or from the Status tab:
Print Monitor
Configuration