6-13
19. Touch the [10.00g/L] button to set the sample value that corresponds to the full scale Voltage.
6.5.5.5 Control Type
20. See Section 6.8 Control Setup for setting up the control function.
21. Touch [Save] to save your Probe A changes and return to the Edit Chamber 1 screen.
22. To monitor an additional chemistry, touch the Probe B [Edit] button.
23. Enter the Chemistry and Units for Probe B. Analog output and Control can also be set for Probe B.
24. Touch [Save] to save your Probe B changes and return to the Edit Chamber 1 screen.
Select Probe
B