5-10
Run Stat
A Stat sample at Station 2 runs without stopping a plate analysis or monitor session that is in progress.
NOTE: The Sipper is not designed to pierce septa.
1. Touch the Run Stat tab
2. Place the Stat sample in Station 2:
a. Insert your sample tube into the tube holder (Station 2) from below the spring clip
b. Slide it up all the way until it rests below the notch at the top.
The test tube holder accepts tubes sizes up to 16x100mm.
Any container other than this should be sampled manually by holding the sample at Station 2.
For a Syringe sample, wait until you are prompted to present the sample.
Figure 5-5
3. Touch [Configure Stat] to setup a Stat sample or [Configure Syringe] to setup a Syringe sample. For a Syringe
sample, the analyzer will wait and allow time for the user to carefully immerse the tip of the sipper into the sample.
4. Select the chemistries and units for the sample.