5-1
5. Running the Instrument
Perform Daily Operational Checks
To ensure that your 2900 Series is operating properly, perform the Membrane Integrity and Linearity checks on a daily
basis before running samples.
5.1.1 Enzyme Membrane Integrity Test
Use YSI 2363 Potassium Ferrocyanide (FCN) Standard to determine if your enzyme membranes are structurally intact.
1. Pour small amount of FCN standard (1000 mg/dL) in a test tube or multi-well plate
2. Press the Run icon to process it as a sample.
3. From the Run Batch tab, touch [New]
4. Choose from the selection of supported racks and plates.
5. We highly recommend renaming the rack/plate “Daily Checks” by touching its ID to indicate that it contains your
daily check batches.
6. After selecting your plate/rack, touch the [Edit] button
7. Touch the location of each sample for the first batch. Selected locations will be blue.