YSI 2900D handleiding

172 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 23 van 172
4-9
Install Calibrator Solution(s)
Caution: To prevent possible damage due to an electrostatic discharge, do NOT touch the metal
tips of the connectors located at the ends of the bottle leads. Handle only the insulated section of
the connectors.
1. Unscrew and remove the lid from the empty calibrator bottle (Cal1ACal3B as indicated).
IMPORTANT: make every effort to avoid contamination of the lid and level sensor assemblies.
2. Mark the date of installation on the label of the new bottle of YSI calibrator solution.
3. Place the new bottle of calibrator in the bottle rack as indicated on the display.
4. Screw the lid and level sensor assembly onto it.
5. Repeat this process for any additional calibrator bottles (Cal1ACal3B as indicated).
Prime the Fluid System
Please note that it may take from several minutes to more than an hour to initially stabilize the probes when setting up for
the first time.
To prime the fluid system:
1. From the Service screen, touch the [Pumps] tab.